DefinePK

DefinePK hosts the largest index of Pakistani journals, research articles, news headlines, and videos. It also offers chapter-level book search.

Use the external radiotherapy StarTrack detector to study the electron beam profile


Article Information

Title: Use the external radiotherapy StarTrack detector to study the electron beam profile

Authors: Safa Sami Hassan

Journal: Journal of Pakistan Medical Association

HEC Recognition History
Category From To
Y 2024-10-01 2025-12-31
X 2023-07-01 2024-09-30
X 2022-07-01 2023-06-30
X 2021-07-01 2022-06-30
X 2020-07-01 2021-06-30
W 2012-07-01 2020-06-30
X 2011-05-13 2012-06-30
Y 1900-01-01 2005-06-30

Publisher: Pakistan Medical Association.

Country: Pakistan

Year: 2024

Volume: 74

Issue: 10 (Supple-08)

Language: en

DOI: 10.47391/JPMA-BAGH-16-18

Categories

Abstract

Objective: To assess the electron beam dose verification, and to investigate the dependability and stability of the Startrack two-dimensional array.
Method: The is a cross-secional study was conducted from January to April 2021 at the Baghdad Centre for Radiotherapy and Nuclear Medicine, Baghdad Medical City, Baghdad, Iraq. quality assurance measurements were made using a StarTrack two-dimensional detector on an electron beam with an energy of 12MeV at 1.5cm depth for field sizes 6cm×6cm and 14cm×14cm. The detector was positioned 100cm from the source to the surface of the linear accelerator infinity. Testing was done using the International Electrotechnical Commissioning protocol.
Results: There was a significant difference between field sizes 6cmx6cm and 14cmx14cm (-4.97±0.83, -2.01±0.68, respectively) (p<0.05). The beam profile measurements of penumbra, flatness and symmetry were within limits ±2 cm, and ±2 cm, 2%, respectively. The percentage of output dose curve against distance showed instability in crossline for 6cm x 6cm and 14cm x 14cm field size.
Conclusion: The output dose was found to be above the tolerance level, and should be corrected before patient treatment for 12MeV energy electron beam therapy.
Key Words: Electrons, Nuclear Medicine, Radiotherapy,